DEVELOPMENT OF A MULTIPLEX REALTIME PCR ASSAY FOR DETECTING COMMON SEXUALLY TRANSMITTED DISEASES PATHOGENS IN HANOI
Main Article Content
Abstract
Objective: 1) To establish a procedure and fabricate a multiplex realtime PCR kit for the detection of common sexually transmitted bacterial and fungal pathogens at the laboratory scale. 2) Determine the infection rate of selected sexually transmitted pathogens in women undergoing gynecological examinations in Hanoi. Subjects: 9 standard samples from Vircell; 135 samples previously confirmed using the commercial PANA RealTyper™ STD kit (Panagene) and careGeneTM STD-12 (Wellsbio); 200 cervical swab specimens from women with vaginal inflammation visiting the Hanoi Obstetrics and Gynecology Hospital between August 2023 and May 2024. Methods: Laboratory experimental combined with cross-sectional description study. Results: The limit of detection (LOD) of the multiplex realtime PCR kit ranged from 1 to 88,81 copies/µl with an accuracy of 100%. The detection capability of the multiplex Realtime PCR kit for 9 pathogens is comparable to the commercial PANA RealTyper™ STD kit (Panagene) and careGeneTM STD-12 (Wellsbio). Among these pathogens, U. parvum has the highest prevalence (41.5%), followed by C. albicans at 19.5%. The remaining pathogens account for lower proportions ranging from 12.0% to 1.5%. Conclusion: A multiplex realtime PCR kit was successfully developed and validated for the simultaneous detection of 9 sexually transmitted pathogens with sensitivity and specificity equivalent to the PANA RealTyper™ STD kit (Panagene) and careGeneTM STD-12 (Wellsbio) certified CE-IVD; detection limit ranges from 1-88.81 copies/µl.
Article Details
Keywords
multiplex realtime PCR, N. gonorrhoeae, C. trachomatis, T. pallidum, H. ducreyi, M. genitalium, M. hominis, U. urealyticum; U. parvum
References

2. Gaydos CA, Cartwright CP, Colaninno P, et al. Performance of the Abbott RealTime CT/NG for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae. J Clin Microbiol. 2010;48(9):3236-3243.

3. Performance of the cobas CT/NG Test Compared to the Aptima AC2 and Viper CTQ/GCQ. Assays for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae. Journal of Clinical Microbiology (2024).

4. qPCR Probes PrimeTime IDT. Integrated DNA Technologies (2024).

5. Bui HTV, Bui HT, Chu SV, et al (2023). Simultaneous real-time PCR detection of nine prevalent sexually transmitted infections using a predesigned double-quenched TaqMan probe panel. PloS One, 18(3), e0282439.

6. Hu XM, Xu JX, Jiang LX, et al (2019). Design and Evaluation of a Novel Multiplex Real-Time PCR Melting Curve Assay for the Simultaneous Detection of Nine Sexually Transmitted Disease Pathogens in Genitourinary Secretions. Front Cell Infect Microbiol. 9, 382.

7. Campos GB, Lobão TN, Selis NN, et al (2015). Prevalence of Mycoplasma genitalium and Mycoplasma hominis in urogenital tract of Brazilian women. BMC Infect Dis.15(1), 60.

8. Sarıer M (2019). Prevalence of Polymicrobial Infection in Urethritis. J Urol Surg. 6(3),180-183.