VALIDATION QUALITY METHOD FOR ROTUNDINE SULFATE IN ROTUNDINE SULFATE INJECTION BY HIGH PERFOMANCE LIQUID CHROMATOGRAPHY
Main Article Content
Abstract
Objective: To validate the assay method for the quantification of rotundine sulfate in rotundine sulfate injection by high performance liquid chromatography (HPLC) in order to contribute to standardization of product quality. Methods: Quantification of Rotundine sulfate was carried out by HPLC method with chromatographic column C18 (4.6×150 mm, 5 µm), flow rate 1.0 mL/min, PDA detector at 281 nm, sample injection volum: 15μL, mobile phase solvent: acetonitrile: water (70:30). Results: The method ensured suitability, specificity, linearity, repeatability, precision, and limits of detection as specified by ICH. Conclusion: The quantitative method responded the requirements and can be used to quantify rotundine sulfate in injection.
Article Details
Keywords
Rotundine sulfate, injection, ICH
References
2. Li L., Zhan X., Li K., Yang X. (2001), Influence of light and heat on the stability of rotundine sulfate injection, Journal of Pharmaceutical Sciences, 90 (10), pp. 1497–1504.
3. Feng M., Zhu Y., Wen J. (2008), HPLC determination of the related substances Rotundine and Rotundine Sulfate Injection, Chinese Journal of Pharmaceutical Analysis, 7, pp.1180-1183.
4. Yin F., Wang X., Li J. (2010), “Determination of Rotundin Sulfae Injection and Related Compounds by HPLC”, Chinese Journal of Pharmaceutical Affairs, 2, pp. 161-163.
5. International Conference on Harmonisation (2005), ICH Q2 (R1) Validation of analytical procedures: Text and methodology.