VALIDATION OF THE PRECISION AND TRUENESS OF QUANTITATIVE HbA1c TESTS USING THE ADAMS HA-8380V DEVICE AT 360 LE HOAN CLINIC

Trọng Đại Nguyễn1, , Việt Tuấn Nguyễn1, Thị Cúc Lương1, Duy Hoan Mai1, Thị Loan Trần1, Thị Hường Lê1, Xuân Hòa Nguyễn1, Trung Nam Lê1, Thị Hoa Nguyễn1, Thị Hạnh Lê1, Thu Hà Phạm2
1 Phòng khám Đa khoa 360 Lê Hoàn
2 Bệnh viện đa khoa Hà Đông

Main Article Content

Abstract

Objective: To verify the precision and trueness of the HbA1c assay performed using the cation-exchange high-performance liquid chromatography (HPLC) method on the ADAMS HA-8380V analyzer at 360 Le Hoan General Clinic.


Materials and Methods: The study utilized control samples at two concentration levels (QC1 and QC2) from Canterbury Scientific (New Zealand). An experimental study design was employed, following the CLSI EP15-A3 guidelines. Results: The within-laboratory coefficients of variation (CVWL) for QC1 and QC2 were 0.87% and 0.40%, respectively; both values ​​were lower than those declared by the manufacturer. The mean values ​​obtained for QC1 and QC2 were 5.532% and 10.796%, respectively, both falling within the validation range. Based on these results, we confirm that the precision and trueness of the HbA1c assay on the ADAMS HA-8380V analyzer are consistent with the manufacturer’s specifications.

Article Details

References

1. Đặng Thị Ngọc Dung. Tổ Chức và Quản Lý Phòng Xét Nghiệm. Nhà xuất bản Y học, Hà Nội; 2020.
2. Clinical and Laboratory Standards Institute. “User Verification of Precision and Estimation of Bias; Approved Guideline-Third Edition 3rd Edition,” CLSI document EP 15A3, pp. 106, 2014.
3. J. O. Westgard, “Desirable Specifications for Total Error, Imprecision, and Bias, derived from intra-and inter-individual biologic variation,” https://www.westgard.com/biodatabase1.htm, Dec 06 2023.
4. N. Đình Thọ, P. T. Mai, và L. V. Thanh. “Đánh giá độ chụm, độ đúng của xét nghiệm HbA1c được định lượng trên thiết bị tosoh hlc-723g8 tại bệnh viện chợ rẫy”, HIUJS, số p.h đặc biệt, tr 55–62, tháng 5 2024.
5. X. Wu, Y. Chao, Z. Wan, Y. Wang, Y. Ma, P. Ke, X. Wu, J. Xu, J. Zhuang and X. Huang, “A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for HbA(1c) determination,” Biochem Med (Zagreb), Vol. 26, No. 3, pp. 353-364, Oct 15 2016.
6. P. Fernández-Riejos, J. M. Maesa, C. S. Mora, M. de Toro, P. M. Valladares, and C. González-Rodriguez, “Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V,” No. 2352-5517, 2016.
7. G. Altawallbeh, V. F. Makky, A. K. Saenger, J. M. Peters, and A. A. Killeen, “Evaluation of an Ion-Exchange HPLC Device for HbA1c Measurement,” J Appl Lab Med.; 5(4):695-703.
8. Urrechaga, E. (2023). Analytical evaluation and quality assessment of the ARKRAY ADAMS A1c Lite HA-8380V for HbA1c. Scandinavian Journal of Clinical and Laboratory Investigation, 83(5), 358–365.