APPLICATION OF CLSI EP15A3 GUIDELINE FOR VERIFICATION OF NEONATAL PHENYLALANINE MEASUREMENT METHOD IN DRY BLOOD SAMPLES ON PERKIN ELMER’S VICTOR 2D SYSTEM

Huy Đông Nguyễn, Thị Kiều Oanh Nguyễn, Thị Huyền Nguyễn, Đình Minh Phạm, Hoàng Bích Nga Lê

Main Article Content

Abstract

Background: Analytical method verification is crucial for laboratories before proceeding to the testing of patients' samples. Objectives: The purpose of this study was to verify the performance of neonatal phenylalanine measurement in dry blood samples on Perkin Elmer’s VICTOR 2D system for newborn screening of Phenylketonuria. Subjects and methods: The verification of precision and trueness was conducted on the Perkin Elmer analyzer using the Perkin Elmer's Neonatal Phenylalanine kit and quality control materials in accordance with CLSI EP15-A3 guidelines. Results: The laboratory CVs of two QC levels were observed to be 18.2% and 8.58% respectively, lower than the manufacturer’s CVs of normal value 33.2% and Abnormal value 14.3%. The observed mean value of two QC levels were within the verification intervals. Conclusion: As a result, the precision and trueness of this Phenylalanine assay were verified, allowing it to be used for neonatal screening.

Article Details

References

1. Westgard J. (2008). Basic Method Validation. 3rd, Westgard QC, Inc., 221–240.
2. International Organization for Standardization (ISO) (2022), Medical laboratories- Requirements for quality and competence.
3. Carey R.N. (2014), User verification of precision and estimation of bias: approved guideline, Clinical and Laboratory Standards Institute, Wayne, Penn.
4. Shoraka H.R., Haghdoost A.A., Baneshi M.R., et al. (2020). Global prevalence of classic phenylketonuria based on Neonatal Screening Program Data: systematic review and meta-analysis. Clin Exp Pediatr, 63(2), 34–43.
5. Minh N.T.B., Long P.T., Khôi L.M., et al. (2021). Nghiên cứu khảo sát tỉ lệ bất thường của xét nghiệm sàng lọc sơ sinh tại Bệnh viện Đại học Y dược thành phố Hồ Chí Minh.
6. Nguyen T., Le Q., Hoang D.T., et al. (2022). Massively parallel sequencing uncovered disease‐associated variant spectra of glucose‐6‐phosphate dehydrogenase deficiency, phenylketonuria and galactosemia in Vietnamese pregnant women. Molec Gen & Gen Med, 10(7), e1959.